Identity

Confirmation that the vial contains the named compound. Included in Silver, Gold and Elite.

At a glance

3 to 5 business days, with purity and quantity.

Included in all levels. LC-MS identity quoted on request.

No additional vial.

ISO 17025 quality system, scope extension in progress

Scope

  • Match of the main peak to the reference standard for the named compound.

Method

  • Retention time and UV spectrum match against the reference standard, from the purity run.
  • LC-MS mass confirmation where chromatography cannot resolve close analogs, or on request. Orbitrap high resolution mass spectrometry where required.

Reported as

  • Match or no match, with the basis stated.
  • Observed mass and spectrum when LC-MS is run.

Sample

  • No additional vial.

Accreditation

Performed under the laboratory’s ISO/IEC 17025:2017 quality system. A2LA scope extension (certificate 6377.01) in progress.

Questions

When is LC-MS used?

For compounds whose analogs cannot be separated by retention time and UV spectrum, and on any order where it is requested.